Conditions of use for multi ball interception impulse voltage generators

Mondo Technology Updated on 2024-02-21

Conditions for the use of multi-ball interception:

Altitude: The system is suitable for areas with an altitude of no more than 1500m.

Ambient Temperature: The normal operating ambient temperature range of the system is -15 to +50.

Relative humidity of the air: The relative humidity of the operating environment of the system should be less than or equal to 90%.

Installation location: The system should be used indoors and is designed to be mobile to facilitate testing in different locations.

Grounding requirements: There must be a reliable grounding point, and the grounding resistance should be less than 1 to ensure the safe and stable operation of the system.

Main technical parameters:

Nominal lightning wave impulse voltage: EDHG-SGSC-200kV, which means that the standard lightning wave impulse voltage that the system can generate is 200kV.

Nominal capacity (energy): 10kJ, which means that the nominal energy output of the system is 10 kJ.

Level capacitance: 20 F, 50kV, means that the cascade capacitance value of the system is 20 microfarads, rated voltage of 50kV.

Level voltage: 50kV, which means that the rated voltage of each stage is 50kV.

Level capacity: 4 25kj, which means that the system consists of 4 levels, and the capacity of each stage is 25 kJ.

Output waveform: The standard lightning impulse full-wave voltage waveform that the system can generate under a load capacitance of 3500pf (including the voltage divider). Specific waveform parameters include wavehead time, wave tail time, and overshoot.

Synchronization range: greater than 20%, indicating that the synchronization operation range of the system exceeds 20%.

Duration of use: The system can work continuously at less than 80% of the rated working voltage; At operating voltages above 80%, the system should operate intermittently to avoid overheating or other damage.

Amplitude adjustment error: less than 1%, which means that the error of the system is controlled within 1% when adjusting the output voltage.

Minimum output voltage: not more than 10% of the nominal voltage of the equipment, that is, the minimum output voltage does not exceed 20kV.

Synchronous misstrike rate: less than 1%, which means that the synchronous operation miserror rate of the system is very low, which ensures the accuracy and reliability of the test.

Base & Size: The base size is 25m × 1.6m, height about 15 meters and weighs about 460kg. The base is equipped with casters for easy movement.

The multi-ball interception impulse voltage generator test system is suitable for the full-wave test of lightning impulse voltage of power products, especially in power products of 10kV and below. It has good environmental adaptability, stable performance and high test accuracy, and is one of the important test equipment in the power system.