Electronic gastroscopic imaging is blurred.
Breakdown repairs. The reason for the blurred imaging of electronic gastroscope is that there is no image, sometimes there is no image, there is no color, and the fog in the center of the image is caused by the desoldering of the CCD component wire, virtual welding, broken wire, short circuit, or micro leakage of pliers and pipes, aging of sealant, etc., and these faults can only be solved by overhaul. However, the only effective way to find these problems in time in clinical use is to carry out leak detection and inspection of the gastroscope before and after each use, and stop using it immediately for maintenance once it is found that there is a micro-leakage.
A video gastroscope is a sophisticated medical device that requires special care and maintenance to ensure its proper function and long-lasting performance. Here are some common maintenance methods for electronic gastroscopes:
Clean and disinfect: During and after use, the electronic gastroscope must be thoroughly cleaned and disinfected. Use appropriate cleaners and disinfectants and follow the manufacturer's guidelines for cleaning and disinfecting.
Pay attention to waterproofing: The electronic gastroscope is a water-sealed device, but special care is still required when cleaning to ensure that water does not enter the inside of the device. Avoid spraying water directly onto the joints or connections, and use a damp cloth to gently wipe the outside.
Store properly: Store the electronic gastroscope in a dry, clean, dust-free environment, away from any liquids or chemicals that may cause damage. It is best to place it in a closed box or shelf to keep dust and debris out.
Regular inspection and maintenance: Regularly inspect and maintain the electronic gastroscope according to the recommendations provided by the manufacturer. This may involve replacing parts, calibrating instruments, or performing other necessary repairs and maintenance.
Avoid collisions and vibrations: Be careful when moving or storing the electronic gastroscope to avoid collisions or strong vibrations to avoid damage to its internal parts.
Replace accessories in a timely mannerReplace the accessories of the electronic gastroscope in a timely manner, such as bulbs and lenses, as needed to ensure the performance and imaging quality of the equipment.
Training and use: Ensure that users are properly trained and instructed to ensure they know how to properly operate and maintain the electronic gastroscope.
Equipment model: Olympus GIF-HQ290 electronic gastroscope.
Test Result:1There are images;
2.The brightness of the beam is acceptable;
3.Bite marks on the insertion tube;
4.Pliers tube wear;
5.The angle is insufficient, the angle is particularly tight, and the snake bone, steel wire, and spiral tube are seriously worn;
Maintenance plan: repair and replace insertion tubes, pliers tubes, snake bones, spiral tubes, steel wires, and erasers.
Maintenance cycle: 5 working days.
Warranty duration: 6 months.