1. The dongle lock is abnormal.
1.The specified dongle 2The clock expires
Cause analysis: Whether the matching dongle is plugged in Whether the USB port of the computer system is good The use time of the factory set dongle expires.
Troubleshooting and solutions: Check whether the red dongle is plugged into the USB port Check whether the indicator light at the tail of the dongle is solid Use a different USB port Upgrade package to upgrade the dongle.
2. The software port is abnormal.
1.Tip: Device 2 not foundThe camera cannot be turned on, and the ** area is shown black.
Cause analysis: Poor contact of serial port line or abnormal drive Abnormal communication of USB port on the computer CCD driver needs to be reset due to signal transmission Unsuccessful installation of CCD driver The operating system and software do not match.
Troubleshooting and Solution: Reinstall the serial cable driver, or plug and unplug the serial cable again. Look at the lower right corner of the screen if there is a prompt installation successful, you can try to replace the serial port Confirm whether the USB hardware port and system communication protocol on the computer are normal Replug and unplug the black camera cable, confirm the reset, and see if there is a prompt installation in the lower right corner of the screen The camera is connected to the host, and reinstall the CCD driver Operating software 32-bit 64-bit 86-bit, configure 32-bit 64-bit 86-bit operating system.
3. Abnormal injection system.
1.Reminder that the water volume is full 2Indicates that the amount of water is insufficient
3.Click on the drops, no drops.
Cause analysis: The rated capacity has been reached, the rated capacity has not been reached, the motor is reminded to form a control out-of-step, the water needle has been placed for too long, and the water evaporation.
Investigation and solution: If the water volume is full, mark the water volume as completed; If the water volume is insufficient, increase the water volume according to the prompt Due to manual operation errors or other accidents, the fixed stroke of the "injection unit" of the instrument is changed, resulting in the instrument injection unit being out of step. Readjust the corresponding stroke of the motor to reset the multiple drops of liquid several times and expel the air until the droplets are discharged.
Fourth, ** display abnormal.
1.Turn on the power to make the injection needle or droplet that cannot see clearly or blur the screen2The dripping needle or water droplet screen appears smeared or stuttered.
Cause analysis: The backlight line is too dim or not turned on The backlight components are damaged The camera magnification adjustment is faulty and not in focus The computer model configuration is relatively low (especially the notebook) The data transfer speed of the USB port is relatively slow The CCD camera software is an old version of the software.
Troubleshooting and solution: Pay attention to whether the light-emitting backlight is on, if it is not bright, because the light is too dim, it will cause a black screen The backlight is on, which proves that the connection is normal. If it doesn't light up, it may be that the internal electrical components are damaged Improve the computer configuration Replace 30 or more USB communication port Upgrade the latest CCD software, or reinstall the camera patch.